Rat Procollagen III N-Terminal Propeptide (PIIINP) ELISA Kit
The Rat Procollagen III N-Terminal Propeptide (PIIINP) ELISA Kit is a powerful tool for measuring levels of PIIINP in rat serum, plasma, and cell culture supernatants with high accuracy and precision. This kit offers exceptional sensitivity and specificity, ensuring dependable and consistent results for various research applications.PIIINP is a key marker for collagen synthesis, specifically for type III collagen, which is essential for connective tissue formation and wound healing.
Elevated levels of PIIINP have been associated with fibrosis, liver disease, and other pathological conditions, making it a valuable biomarker for investigating these diseases and potential therapeutic interventions.Overall, the Rat Procollagen III N-Terminal Propeptide (PIIINP) ELISA Kit is a reliable and effective tool for researchers studying collagen metabolism, fibrosis, and related disorders in rat models.
Product Name:
Rat Procollagen III N-Terminal Propeptide (PIIINP) ELISA Kit
Product Code:
RTDL00830
Size:
96 Assays
Target:
PIIINP
Synonyms:
P3NP, N-Propeptide Of Type III Procollagen, Procollagen III Amino Terminal Propeptide
Detection Method:
Sandwich
Range:
31.2-2,000pg/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.