Rat Wingless Type MMTV Integration Site Family, Member 4 (WNT4) ELISA Kit
The Rat Wnt4 (Wingless-type MMTV Integration site family member 4) ELISA Kit is a valuable tool for accurately measuring levels of Wnt4 in rat serum, plasma, and cell culture supernatants. This kit boasts high sensitivity and specificity, ensuring precise and reliable results for a variety of research applications.Wnt4 is a critical protein involved in cellular signaling pathways, particularly in development and tissue regeneration.
Its dysregulation has been implicated in various diseases, including cancer, metabolic disorders, and developmental abnormalities, making it a key biomarker for understanding these conditions and exploring potential therapeutic interventions.Overall, the Rat Wnt4 ELISA Kit offers researchers a comprehensive and efficient solution for investigating the roles of Wnt4 in biological processes and disease mechanisms.
Product Name:
Rat Wingless Type MMTV Integration Site Family, Member 4 (WNT4) ELISA Kit
Product Code:
RTDL01117
Size:
96 Assays
Target:
WNT4
Detection Method:
Sandwich
Range:
0.156-10ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.