Rat Trypsinogen Activation Peptide (TAP) ELISA Kit
The Rat Trypsinogen Activation Peptide (TAP) ELISA Kit is a powerful tool for studying trypsinogen activation in rat samples. This kit is designed for the accurate detection of TAP levels in rat serum, plasma, and cell culture supernatants with high sensitivity and specificity. With reliable and reproducible results, researchers can confidently use this kit for a wide range of research applications.TAP is a key peptide involved in trypsinogen activation, a process essential for proper digestive enzyme regulation.
Dysregulation of trypsinogen activation has been implicated in various gastrointestinal disorders, making TAP a valuable biomarker for studying these conditions and developing potential therapies.Overall, the Rat TAP ELISA Kit offers researchers a reliable and efficient method for studying trypsinogen activation in rat samples, providing valuable insights into digestive enzyme regulation and associated disorders.
Product Name:
Rat Trypsinogen Activation Peptide (TAP) ELISA Kit
Product Code:
RTDL01004
Size:
96 Assays
Target:
TAP
Detection Method:
Competition
Range:
0.156-10ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-99
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition. Note:To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity, and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 50µL standard or sample to each well.    And then add 50µL prepared Detection Reagent A immediately.    Shake and mix. Incubate 1 hour at 37°C
3.
Aspirate and wash 3 times
4.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
5.
Aspirate and wash 5 times
6.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
7.
Add 50µL Stop Solution. Read at 450 nm immediately.