The Rat Thymic Stromal Lymphopoietin (TSLP) ELISA Kit is specially designed for the precise detection of TSLP levels in rat serum, plasma, and tissue homogenates. With its high sensitivity and specificity, this kit offers reliable and reproducible results, making it an excellent tool for various research applications.TSLP is a key cytokine that plays a crucial role in regulating immune responses and is involved in the development of allergic inflammation and autoimmune diseases.
Its measurement is essential for understanding the mechanisms of these conditions and for the development of targeted therapies.This ELISA kit is a valuable resource for researchers studying immune system modulation, allergic diseases, and inflammatory disorders in rats. It offers a simple and efficient way to quantify TSLP levels, providing valuable insights into its role in health and disease.
Product Name:
Rat Thymic Stromal Lymphopoietin (TSLP) ELISA Kit
Product Code:
RTDL01070
Size:
96 Assays
Target:
TSLP
Detection Method:
Sandwich
Range:
15.625-1000pg/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.