The Rat Relaxin (RLN) ELISA Kit is a powerful tool for measuring relaxin levels in rat serum, plasma, and cell culture supernatants. With its high sensitivity and specificity, this kit provides accurate and reliable results for a variety of research applications.Relaxin is a hormone that plays a key role in regulating various physiological processes, including reproductive function and cardiovascular health. It is also involved in processes such as tissue remodeling and inflammation.
By accurately measuring relaxin levels, researchers can gain insights into its role in different diseases and conditions, leading to potential therapeutic advancements.Whether studying reproductive physiology, cardiovascular diseases, or inflammatory disorders, the Rat Relaxin (RLN) ELISA Kit provides a valuable tool for researchers looking to understand the role of relaxin in health and disease.
Product Name:
Rat Relaxin (RLN) ELISA Kit
Product Code:
RTDL00908
Size:
96 Assays
Target:
RLN
Synonyms:
RLXH1, Relaxin H1, Prorelaxin H1
Detection Method:
Competition
Range:
31.2-2000pg/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-99
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition. Note:To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity, and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 50µL standard or sample to each well.    And then add 50µL prepared Detection Reagent A immediately.    Shake and mix. Incubate 1 hour at 37°C
3.
Aspirate and wash 3 times
4.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
5.
Aspirate and wash 5 times
6.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
7.
Add 50µL Stop Solution. Read at 450 nm immediately.