Rat Receptor Activator Of Nuclear Factor Kappa B Ligand (RANkL) ELISA Kit
The Rat Receptor Activator of Nuclear Factor Kappa B Ligand (RANKL) ELISA Kit is a highly specialized assay designed for the precise measurement of RANKL levels in rat serum, plasma, and tissue culture samples. This kit offers exceptional sensitivity and specificity, ensuring accurate and consistent results for a variety of research applications.RANKL is a key regulator of bone metabolism and immune response, playing a critical role in osteoclast differentiation and activation. Dysregulation of RANKL signaling has been implicated in various bone diseases, including osteoporosis and rheumatoid arthritis, making it a valuable biomarker for studying these conditions and developing targeted therapies.
With the Rat RANKL ELISA Kit, researchers can confidently quantify RANKL levels in rat samples with confidence and precision, advancing their understanding of bone physiology and disease pathology. Trust in the reliability and performance of this specialized assay for your next research study.
Product Name:
Rat Receptor Activator Of Nuclear Factor Kappa B Ligand (RANkL) ELISA Kit
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
87-100
93
EDTA plasma (n=5)
89-99
94
Heparin plasma (n=5)
90-98
94
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.