Rat Plasminogen Activator Inhibitor 2 (PAI2) ELISA Kit
The Rat Plasminogen Activator Inhibitor-2 (PAI-2) ELISA Kit is a powerful tool for measuring levels of PAI-2 in rat samples with high accuracy and precision. This kit is suitable for use with rat serum, plasma, and cell culture supernatants, and offers exceptional sensitivity and specificity for consistent and dependable results.PAI-2 is a key regulator of the fibrinolytic system, playing a crucial role in controlling blood clot formation and dissolution. Dysregulation of PAI-2 has been implicated in various diseases, including cardiovascular disorders, cancer, and inflammatory conditions.
Therefore, this ELISA kit serves as a valuable resource for researchers studying these pathologies and exploring potential therapeutic interventions.Overall, the Rat PAI-2 ELISA Kit from Assay Genie provides researchers with a reliable and efficient tool for investigating the role of PAI-2 in disease progression and may ultimately contribute to the development of novel treatment strategies.
Product Name:
Rat Plasminogen Activator Inhibitor 2 (PAI2) ELISA Kit
Product Code:
RTDL00796
Size:
96 Assays
Target:
PAI2
Synonyms:
SERPINB2, PLANH2, Serpin Peptidase Inhibitor Clade B Member 2, Monocyte Arg-serpin, Placental plasminogen activator inhibitor, Serpin B2, Urokinase inhibitor
Detection Method:
Sandwich
Range:
1.56-100ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.