The Rat Enolase Non-Neuronal (NNE) ELISA Kit is a powerful tool for the quantitative measurement of rat NNE levels in biological samples such as serum, plasma, and cell culture supernatants. This kit has been carefully designed to provide accurate and reliable results, with high sensitivity and specificity to ensure reproducibility.Enolase is an enzyme that plays a crucial role in glycolysis, the metabolic process that produces energy in cells. The non-neuronal isoform of enolase is found in various tissues and is involved in multiple biological processes. Abnormal levels of NNE have been associated with different diseases, including cancer, cardiovascular disorders, and neurodegenerative conditions, making it a key biomarker for research and potential therapeutic development.
By using the Rat Enolase Non-Neuronal ELISA Kit, researchers can gain valuable insights into the role of NNE in health and disease, paving the way for advancements in understanding and treating these conditions. With its superior performance and accuracy, this kit is an essential tool for studying the physiological and pathological functions of enolase in rat models.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-99
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.