Rat Corticotropin Releasing Hormone (CRH) ELISA Kit
The Rat Corticotropin-Releasing Hormone (CRH) ELISA Kit is a powerful tool for measuring CRH levels in rat samples such as serum, plasma, and tissue lysates. With its high sensitivity and specificity, this kit provides accurate and reliable results for a variety of research applications.CRH is a key regulator of the body's response to stress and plays a crucial role in the neuroendocrine system. Dysregulation of CRH has been implicated in various stress-related disorders such as anxiety, depression, and PTSD.
By quantifying CRH levels, researchers can gain valuable insights into the physiological and pathological processes involving this hormone.Overall, the Rat CRH ELISA Kit is essential for studying stress-related disorders, developing potential treatments, and advancing our understanding of the neuroendocrine system. Trust AssayGenie for high-quality ELISA kits for your research needs.
Product Name:
Rat Corticotropin Releasing Hormone (CRH) ELISA Kit
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-99
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition. Note:To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity, and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 50µL standard or sample to each well.    And then add 50µL prepared Detection Reagent A immediately.    Shake and mix. Incubate 1 hour at 37°C
3.
Aspirate and wash 3 times
4.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
5.
Aspirate and wash 5 times
6.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
7.
Add 50µL Stop Solution. Read at 450 nm immediately.