The Rat Cathepsin B (CTSB) ELISA Kit is specially designed for the precise measurement of rat Cathepsin B levels in biological samples such as serum, plasma, and tissue homogenates. With its high sensitivity and specificity, this kit ensures accurate and reproducible results, making it an invaluable tool for a variety of research applications.Cathepsin B is a key enzyme involved in cellular processes such as protein degradation and tissue remodeling. Abnormal levels of Cathepsin B have been linked to various diseases including cancer, cardiovascular disorders, and inflammatory conditions, making it a promising biomarker for understanding disease pathogenesis and developing targeted therapies.
With the Rat Cathepsin B (CTSB) ELISA Kit, researchers can confidently investigate the role of Cathepsin B in disease progression and explore its potential as a therapeutic target. Order your kit today to advance your research efforts in the field of molecular biology and disease mechanisms.
Product Name:
Rat Cathepsin B (CTSB) ELISA Kit
Product Code:
RTDL00263
Size:
96 Assays
Target:
CTSB
Synonyms:
CTS-B, APPS, CPSB, APP secretase
Detection Method:
Sandwich
Range:
0.156-10ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.