Rat AR (Amphiregulin) ELISA Kit
- SKU:
- AEES00462
- Product Type:
- ELISA Kit
- ELISA Type:
- Sandwich
- Size:
- 96 Assays
- Reactivity:
- Rat
- Sensitivity:
- 18.75 pg/mL
- Range:
- 31.25-2000 pg/mL
- Sample Type:
- Serum, plasma and other biological fluids
Description
Rat AR (Amphiregulin) ELISA Kit
The Rat AR (Amphiregulin) ELISA Kit is a specialized assay tool designed for the precise quantification of Amphiregulin levels in various biological samples obtained from rat models.
Amphiregulin is a member of the epidermal growth factor (EGF) family, known for its crucial regulatory roles in promoting cell growth, proliferation, differentiation, and migration. Its functions extend to diverse biological processes and mechanisms within the body. This ELISA kit, crafted by Assay Genie, ensures exceptional sensitivity and specificity, enabling researchers to obtain accurate and reproducible results. With robust performance and ease of use, this kit is manufactured under stringent quality control measures, making it an ideal choice for research studies focused on understanding the biological activities of Amphiregulin in rats and its implications in various physiological processes.
Product Name: | Rat AR (Amphiregulin) ELISA Kit |
Product Code: | AEES00462 |
Assay Type: | Sandwich |
Format: | 96T |
Assay Time: | 3.5h |
Reactivity: | Rat |
Detection Range: | 31.25-2000 pg/mL |
Sensitivity: | 18.75 pg/mL |
Sample Type & Sample Volume: | Serum, plasma and other biological fluids, 100μL |
Specificity: | This kit recognizes Rat AR in samples. No significant cross-reactivity or interference between Rat AR and analogues was observed. |
Reproducibility: | Both intra-CV and inter-CV are < 10%. |
Application: | This ELISA kit applies to the in vitro quantitative determination of Rat AR concentrations in Serum, plasma and other biological fluids. |
This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Rat AR. Standards or samples are added to the micro ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Rat AR and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Rat AR, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The OD value is proportional to the concentration of Rat AR. You can calculate the concentration of Rat AR in the samples by comparing the OD of the samples to the standard curve.
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level Rat AR were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level Rat AR were tested on 3 different plates, 20 replicates in each plate.
Intra-assay Precision | Inter-assay Precision | |||||
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 104.99 | 236.79 | 752.86 | 108.88 | 253.87 | 690.95 |
Standard deviation | 6.59 | 13.17 | 23.79 | 6.53 | 15.03 | 37.17 |
CV (%) | 6.28 | 5.56 | 3.16 | 6.00 | 5.92 | 5.38 |
Recovery
The recovery of Rat AR spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type | Range (%) | Average Recovery (%) |
Serum(n=8) | 88-99 | 94 |
EDTA plasma (n=8) | 84-96 | 91 |
Cell culture media (n=8) | 92-107 | 98 |
Kit Components: | An unopened kit can be stored at 2-8℃ for 1 month. If the kit is not supposed to be used within 1 month, store the components separately according to the following conditions once the kit is received.
|
1. | Add 100μL standard or sample to the wells. Incubate for 90 min at 37°C |
2. | Discard the liquid, immediately add 100μL Biotinylated Detection Ab working solution to each well. Incubate for 60 min at 37°C |
3. | Aspirate and wash the plate for 3 times |
4. | Add 100μL HRP conjugate working solution. Incubate for 30 min at 37°C. Aspirate and wash the plate for 5 times |
5. | Add 90μL Substrate Reagent. Incubate for 15 min at 37°C |
6. | Add 50μL Stop Solution |
7. | Read the plate at 450nm immediately. Calculation of the results |