The Rat Amyloid Beta Peptide 1-40 (AB1-40) ELISA Kit is a powerful tool for the accurate measurement of amyloid beta peptide levels in rat samples, including serum, plasma, and cell culture supernatants. With its high sensitivity and specificity, this kit delivers reliable and reproducible results for a variety of research applications.Amyloid beta peptides play a key role in the pathogenesis of Alzheimer's disease, making them a critical target for studying neurodegenerative disorders.
By accurately quantifying AB1-40 levels, researchers can gain valuable insights into the progression of these conditions and potentially identify new therapeutic approaches.Overall, the Rat Amyloid Beta Peptide 1-40 ELISA Kit offers researchers a valuable tool for investigating the role of amyloid beta peptides in neurodegenerative diseases and advancing our understanding of potential treatment options.
Product Name:
Rat Amyloid Beta Peptide 1-40 (Ab1-40) ELISA Kit
Product Code:
RTDL00004
Size:
96 Assays
Target:
Ab1-40
Detection Method:
Competition
Range:
31.25-2000pg/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
83-100
91
EDTA plasma (n=5)
91-98
94
Heparin plasma (n=5)
90-99
94
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition. Note:To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity, and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 50µL standard or sample to each well.    And then add 50µL prepared Detection Reagent A immediately.    Shake and mix. Incubate 1 hour at 37°C
3.
Aspirate and wash 3 times
4.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
5.
Aspirate and wash 5 times
6.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
7.