The Rabbit Tissue Factor Pathway Inhibitor (TFPI) ELISA Kit is a powerful tool for researchers looking to detect levels of TFPI in rabbit serum, plasma, and tissue samples. This kit offers exceptional sensitivity and specificity, ensuring accurate and consistent results for a variety of research applications.TFPI is a key regulator of the blood clotting pathway, specifically inhibiting the tissue factor pathway. Dysregulation of TFPI has been implicated in various diseases and conditions, including thrombosis, cardiovascular diseases, and inflammatory disorders.
Monitoring TFPI levels can provide valuable insights into these conditions and may lead to the development of new treatment strategies.With its user-friendly protocol and reliable performance, the Rabbit TFPI ELISA Kit is an essential tool for researchers investigating the role of TFPI in health and disease. Stay ahead of the curve with this cutting-edge kit and uncover the secrets of the tissue factor pathway inhibitor in rabbit models.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.