The Rabbit Interleukin-1 Beta (IL1B) ELISA Kit is a powerful tool for detecting and measuring levels of interleukin-1 beta in rabbit serum, plasma, and cell culture supernatants. This kit offers exceptional sensitivity and specificity, ensuring accurate and reliable results for a variety of research applications.Interleukin-1 beta is a critical cytokine involved in the inflammatory response, playing a key role in various physiological and pathological processes. Its dysregulation has been linked to numerous inflammatory conditions, autoimmune diseases, and cancer.
By accurately measuring IL1B levels, researchers can gain valuable insights into disease mechanisms and potential therapeutic targets.With its user-friendly protocol and robust performance, the Rabbit Interleukin-1 Beta (IL1B) ELISA Kit is an essential tool for studying the role of IL1B in health and disease, advancing research in immunology, inflammation, and beyond.
Product Name:
Rabbit Interleukin 1 Beta (IL-1b) ELISA Kit
Product Code:
RBDL00034
Size:
96 Assays
Target:
IL1b
Synonyms:
IL1-B, IL1-Beta, IL1F2, IL-1?, Interleukin-1 Family Member 2, Catabolin
Detection Method:
Sandwich
Range:
78.125-5000pg/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
83-100
91
EDTA plasma (n=5)
90-103
96
Heparin plasma (n=5)
89-102
95
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.