The Porcine Vascular Cell Adhesion Molecule-1 (VCAM-1) ELISA Kit is a highly reliable and sensitive tool for the precise measurement of VCAM-1 levels in porcine serum, plasma, and cell culture supernatants. This kit offers exceptional specificity and accuracy, guaranteeing consistent and dependable results for various research applications.VCAM-1 is a critical molecule involved in vascular inflammation and cell adhesion, playing a crucial role in a variety of biological processes. Its dysregulation has been associated with various diseases such as atherosclerosis, inflammatory disorders, and autoimmune conditions, making it a valuable biomarker for investigating these pathologies and potential therapeutic interventions.
Whether studying vascular biology, immunology, or disease mechanisms, the Porcine VCAM-1 ELISA Kit provides researchers with a powerful tool for comprehensive analysis and exploration of VCAM-1 functions in porcine models. Trust in the quality and precision of this ELISA kit for accurate and insightful results in your studies.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.