Porcine Platelet Derived Growth Factor Subunit B (PDGFB) ELISA Kit
The Porcine Platelet-Derived Growth Factor Subunit B (PDGFB) ELISA Kit is specifically designed for the accurate quantification of PDGFB levels in porcine serum, plasma, and cell culture supernatants. This kit offers exceptional sensitivity and specificity, ensuring precise and reliable results for a variety of research applications.PDGFB is a key growth factor that plays a crucial role in cell growth, division, and angiogenesis. It is involved in various biological processes such as wound healing and tissue regeneration. The accurate measurement of PDGFB levels can provide valuable insights into the pathogenesis of diseases such as cancer, cardiovascular diseases, and inflammatory disorders.
With its high performance and ease of use, the Porcine PDGFB ELISA Kit is a valuable tool for researchers studying the role of PDGFB in different physiological and pathological processes. Its reliability and accuracy make it an essential component of any research laboratory focused on understanding growth factor signaling pathways and developing potential therapeutic interventions.
Product Name:
Porcine Platelet Derived Growth Factor Subunit B (PDGFB) ELISA Kit
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.