The Porcine Matrix Metalloproteinase-3 (MMP3) ELISA Kit is a highly reliable and sensitive assay designed for the accurate measurement of MMP3 levels in porcine serum, plasma, and cell culture supernatants. This kit offers unparalleled specificity, allowing for precise and reproducible results that are essential for various research applications.MMP3, also known as stromelysin-1, is a critical enzyme involved in the breakdown of extracellular matrix proteins and tissue remodeling. Elevated levels of MMP3 have been linked to various pathological conditions, including arthritis, cancer metastasis, and inflammatory diseases.
Therefore, the detection of MMP3 levels can serve as a valuable marker for assessing disease progression and evaluating potential treatment strategies.With its user-friendly protocol and high-quality components, the Porcine MMP3 ELISA Kit is an indispensable tool for researchers studying the role of MMP3 in disease pathogenesis and therapeutic interventions. Trust in this kit to deliver reliable and accurate results that will advance your research efforts in the field of matrix metalloproteinase biology.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-99
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.