The Mouse Synuclein Alpha (SNCA) ELISA Kit is specifically designed for the precise quantification of alpha-synuclein levels in mouse samples, including serum, plasma, and tissue homogenates. With its high sensitivity and specificity, this kit ensures accurate and reproducible results, making it an excellent tool for a variety of research applications.Alpha-synuclein is a key protein involved in the pathogenesis of neurodegenerative disorders, particularly Parkinson's disease.
By measuring alpha-synuclein levels, researchers can gain valuable insights into disease progression and potential therapeutic targets. This ELISA kit provides a reliable method for studying alpha-synuclein dynamics in mouse models, advancing our understanding of neurodegenerative diseases and facilitating the development of new treatment strategies.
Product Name:
Mouse Synuclein Alpha (SNCa) ELISA Kit
Product Code:
MODL01209
Size:
96 Assays
Target:
SNCa
Synonyms:
SNC-A, aSYN, PD1, PARK1, PARK4, NACP, Non-A Beta Component Of Alzheimer's Disease Amyloid Precursor Protein, Non-A4 component of amyloid precursor
Detection Method:
Sandwich
Range:
15.625-1,000pg/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.