Mouse S100 Calcium Binding Protein B (S100B) ELISA Kit
The Mouse S100 Calcium Binding Protein B (S100B) ELISA Kit offered by Assay Genie is specifically designed for the accurate detection of S100B levels in mouse serum, plasma, and cell culture supernatants. This kit boasts high sensitivity and specificity, ensuring reliable and reproducible results for a variety of research applications.S100B is a key protein involved in calcium signaling and regulation, with implications in various biological processes including cell growth, differentiation, and inflammation.
Elevated levels of S100B have been associated with conditions such as neurodegenerative disorders, brain injury, and certain types of cancer, highlighting its importance as a biomarker for studying these diseases and potential therapeutic interventions.With its advanced design and precision, the Mouse S100 Calcium Binding Protein B (S100B) ELISA Kit from Assay Genie is a valuable tool for researchers looking to explore the role of S100B in health and disease.
Product Name:
Mouse S100 Calcium Binding Protein B (S100B) ELISA Kit
Product Code:
MODL01147
Size:
96 Assays
Target:
S100B
Synonyms:
S100-B, NEF, S100Beta, S-100 protein subunit beta
Detection Method:
Sandwich
Range:
15.6-1,000pg/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
83-100
91
EDTA plasma (n=5)
91-102
96
Heparin plasma (n=5)
89-102
95
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.