Mouse S100 Calcium Binding Protein A10 (S100A10) ELISA Kit
The Mouse S100 Calcium Binding Protein A10 (S100A10) ELISA Kit is a highly sensitive and specific assay designed for the accurate measurement of S100A10 levels in mouse serum, plasma, and cell culture supernatants. This kit provides reliable and reproducible results, making it an excellent tool for a variety of research applications.S100A10, also known as p11, is a calcium-binding protein that plays a crucial role in various cellular processes, including cell cycle regulation, cell adhesion, and membrane trafficking.
Dysregulation of S100A10 has been implicated in diseases such as cancer, inflammation, and neurodegenerative disorders, making it a valuable biomarker for studying these conditions and potential therapeutic targets.Overall, the Mouse S100A10 ELISA Kit offers researchers a powerful tool for investigating the role of S100A10 in health and disease, providing valuable insights into its mechanisms and potential therapeutic applications.
Product Name:
Mouse S100 Calcium Binding Protein A10 (S100A10) ELISA Kit
Product Code:
MODL01141
Size:
96 Assays
Target:
S100A10
Synonyms:
ANX2L, ANX2LG, CAL1L, CLP11, GP11, P11, P10, Calpactin-1 light chain, Cellular ligand of annexin II, Annexin II Ligand,Calpactin I,Light Polypeptide
Detection Method:
Sandwich
Range:
0.312-20ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
83-100
91
EDTA plasma (n=5)
90-98
94
Heparin plasma (n=5)
89-102
95
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.