Mouse PECAM1 (Platelet/Endothelial Cell Adhesion Molecule 1) ELISA Kit (MOES01377)
- SKU:
- MOES01377
- Product Type:
- ELISA Kit
- Size:
- 96 Assays
- Uniprot:
- Q08481
- Sensitivity:
- 9.38pg/mL
- Range:
- 15.63-1000pg/mL
- ELISA Type:
- Sandwich
- Reactivity:
- Mouse
- Sample Type:
- Serum, plasma and other biological fluids
- Research Area:
- Cell Biology
Description
Mouse PECAM1 (Platelet/Endothelial Cell Adhesion Molecule 1) ELISA Kit
The Mouse PECAM-1 (Platelet Endothelial Cell Adhesion Molecule-1) ELISA Kit is a powerful tool for researchers looking to accurately measure PECAM-1 levels in mouse samples. This kit offers exceptional sensitivity and specificity, providing reliable and consistent results for a variety of research applications.PECAM-1 is a key protein involved in cell adhesion and plays a crucial role in endothelial cell functions, immune responses, and inflammation. Its dysregulation has been linked to various diseases, including cardiovascular disorders, inflammatory conditions, and cancer.
Therefore, measuring PECAM-1 levels can provide valuable insights into the pathogenesis of these diseases and aid in the development of targeted therapies.With the Mouse PECAM-1 ELISA Kit from Assay Genie, researchers can accurately quantify PECAM-1 levels in mouse serum, plasma, and tissue culture supernatants, allowing for in-depth investigation of its biological functions and potential therapeutic implications.
Assay type: | Sandwich |
Format: | 96T |
Assay time: | 4.5h |
Reactivity: | Mouse |
Detection Method: | Colormetric |
Detection Range: | 15.63-1000 pg/mL |
Sensitivity: | 9.38 pg/mL |
Sample Volume Required Per Well: | 100µL |
Sample Type: | Serum, plasma and other biological fluids |
Specificity: | This kit recognizes Mouse PECAM1 in samples. No significant cross-reactivity or interference between Mouse PECAM1 and analogues was observed. |
This ELISA kit uses Sandwich-ELISA as the method. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Mouse PECAM1. Standards or samples are added to the appropriate micro ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Mouse PECAM1 and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well successively and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Mouse PECAM1, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by adding Stop Solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The OD value is proportional to the concentration of Mouse PECAM1. The concentration of Mouse PECAM1 in samples can be calculated by comparing the OD of the samples to the standard curve.
UniProt Protein Function: | PECAM-1: a cell adhesion molecule expressed on platelets and at endothelial cell intercellular junctions. Plays a role in modulation of integrin-mediated cell adhesion, transendothelial migration, angiogenesis, apoptosis, cell migration, negative regulation of immune cell signaling, autoimmunity, macrophage phagocytosis, IgE-mediated anaphylaxis and thrombosis. Undergoes alternative splicing, generating multiple isoforms in vascular beds of various tissues. |
UniProt Protein Details: | Protein type:Cell adhesion; Membrane protein, integral Cellular Component: extracellular space; cell surface; smooth muscle contractile fiber; membrane; integral to membrane; plasma membrane; intracellular; cell junction; lipid raft; external side of plasma membrane Molecular Function:protein binding Biological Process: negative regulation of peptidyl-tyrosine phosphorylation; wound healing; positive regulation of tyrosine phosphorylation of Stat5 protein; negative regulation of actin filament polymerization; positive regulation of leukocyte migration; diapedesis; cell adhesion; regulation of cell migration; phagocytosis; Rho protein signal transduction |
UniProt Code: | Q08481 |
NCBI GenInfo Identifier: | 585658 |
NCBI Gene ID: | 18613 |
NCBI Accession: | Q08481. 1 |
UniProt Related Accession: | Q08481 |
Molecular Weight: | |
NCBI Full Name: | Platelet endothelial cell adhesion molecule |
NCBI Synonym Full Names: | platelet/endothelial cell adhesion molecule 1 |
NCBI Official Symbol: | Pecam1 |
NCBI Official Synonym Symbols: | Cd31; Pecam; C85791; PECAM-1 |
NCBI Protein Information: | platelet endothelial cell adhesion molecule |
UniProt Protein Name: | Platelet endothelial cell adhesion molecule |
UniProt Synonym Protein Names: | CD_antigen: CD31 |
UniProt Gene Name: | Pecam1 |
UniProt Entry Name: | PECA1_MOUSE |
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e. g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration (pg/mL) | O.D | Average | Corrected |
1000 | 2.471 2.511 | 2.491 | 2.41 |
500 | 1.663 1.679 | 1.671 | 1.59 |
250 | 1.002 0.986 | 0.994 | 0.913 |
125 | 0.494 0.514 | 0.504 | 0.423 |
62.5 | 0.255 0.247 | 0.251 | 0.17 |
31.25 | 0.196 0.168 | 0.182 | 0.101 |
15.63 | 0.126 0.14 | 0.133 | 0.052 |
0 | 0.074 0.088 | 0.081 | -- |
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level Mouse PECAM1 were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level Mouse PECAM1 were tested on 3 different plates, 20 replicates in each plate.
Intra-assay Precision | Inter-assay Precision | |||||
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 45.60 | 116.94 | 370.17 | 49.93 | 116.39 | 347.54 |
Standard deviation | 3.11 | 5.81 | 18.84 | 3.05 | 6.59 | 19.01 |
C V (%) | 6.82 | 4.97 | 5.09 | 6.11 | 5.66 | 5.47 |
Recovery
The recovery of Mouse PECAM1 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type | Range (%) | Average Recovery (%) |
Serum (n=5) | 86-101 | 93 |
EDTA plasma (n=5) | 88-102 | 93 |
Cell culture media (n=5) | 91-106 | 98 |
Linearity
Samples were spiked with high concentrations of Mouse PECAM1 and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) | EDTA plasma (n=5) | Cell culture media (n=5) | ||
1:2 | Range (%) | 89-102 | 86-98 | 87-101 |
Average (%) | 94 | 91 | 94 | |
1:4 | Range (%) | 89-105 | 86-99 | 82-95 |
Average (%) | 96 | 92 | 89 | |
1:8 | Range (%) | 87-102 | 85-98 | 84-94 |
Average (%) | 93 | 91 | 89 | |
1:16 | Range (%) | 88-100 | 85-99 | 82-92 |
Average (%) | 95 | 91 | 87 |
An unopened kit can be stored at 4°C for 1 month. If the kit is not used within 1 month, store the items separately according to the following conditions once the kit is received.
Item | Specifications | Storage |
Micro ELISA Plate(Dismountable) | 8 wells ×12 strips | -20°C, 6 months |
Reference Standard | 2 vials | |
Concentrated Biotinylated Detection Ab (100×) | 1 vial, 120 µL | |
Concentrated HRP Conjugate (100×) | 1 vial, 120 µL | -20°C(shading light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 4°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | |
HRP Conjugate Diluent | 1 vial, 14 mL | |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | |
Substrate Reagent | 1 vial, 10 mL | 4°C(shading light) |
Stop Solution | 1 vial, 10 mL | 4°C |
Plate Sealer | 5 pieces | |
Product Description | 1 copy | |
Certificate of Analysis | 1 copy |
- Set standard, test sample and control (zero) wells on the pre-coated plate and record theirpositions. It is recommended to measure each standard and sample in duplicate. Note: addall solutions to the bottom of the plate wells while avoiding contact with the well walls. Ensuresolutions do not foam when adding to the wells.
- Aliquot 100µl of standard solutions into the standard wells.
- Add 100µl of Sample / Standard dilution buffer into the control (zero) well.
- Add 100µl of properly diluted sample (serum, plasma, tissue homogenates and otherbiological fluids) into test sample wells.
- Cover the plate with the sealer provided in the kit and incubate for 90 min at 37°C.
- Aspirate the liquid from each well, do not wash. Immediately add 100µL of BiotinylatedDetection Ab working solution to each well. Cover the plate with a plate seal and gently mix. Incubate for 1 hour at 37°C.
- Aspirate or decant the solution from the plate and add 350µL of wash buffer to each welland incubate for 1-2 minutes at room temperature. Aspirate the solution from each well andclap the plate on absorbent filter paper to dry. Repeat this process 3 times. Note: a microplatewasher can be used in this step and other wash steps.
- Add 100µL of HRP Conjugate working solution to each well. Cover with a plate seal andincubate for 30 min at 37°C.
- Aspirate or decant the solution from each well. Repeat the wash process for five times asconducted in step 7.
- Add 90µL of Substrate Reagent to each well. Cover with a new plate seal and incubate forapproximately 15 min at 37°C. Protect the plate from light. Note: the reaction time can beshortened or extended according to the actual color change, but not by more than 30min.
- Add 50 µL of Stop Solution to each well. Note: Adding the stop solution should be done inthe same order as the substrate solution.
- Determine the optical density (OD value) of each well immediately with a microplate readerset at 450 nm.