The Mouse Parathyroid Hormone (PTH) ELISA Kit is specifically developed for the precise measurement of parathyroid hormone levels in mouse serum, plasma, and cell culture supernatants. With exceptional sensitivity and specificity, this kit guarantees consistent and accurate results, making it a valuable tool for various research studies.Parathyroid hormone is a vital hormone responsible for regulating calcium and phosphorus levels in the body. Imbalances in parathyroid hormone levels can lead to bone disorders, kidney diseases, and other metabolic issues.
Therefore, monitoring PTH levels is crucial in understanding and treating these conditions effectively.This ELISA kit offers researchers a reliable method for quantifying mouse PTH levels, facilitating in-depth studies on calcium homeostasis, bone metabolism, and related diseases. With its user-friendly protocols and high-quality components, the Mouse PTH ELISA Kit is an indispensable resource for advancing scientific research in the field of endocrinology.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-99
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition. Note:To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity, and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 50µL standard or sample to each well.    And then add 50µL prepared Detection Reagent A immediately.    Shake and mix. Incubate 1 hour at 37°C
3.
Aspirate and wash 3 times
4.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
5.
Aspirate and wash 5 times
6.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
7.
Add 50µL Stop Solution. Read at 450 nm immediately.