The Mouse P-Selectin Glycoprotein Ligand-1 (PSGL-1) ELISA Kit is a reliable and accurate tool for detecting levels of PSGL-1 in mouse serum, plasma, and cell culture supernatants. This kit is highly sensitive and specific, ensuring precise and reproducible results for a variety of research applications.PSGL-1 is a critical protein involved in mediating cell adhesion and migration, playing a pivotal role in inflammation, immune response, and the regulation of blood flow. Studying PSGL-1 levels can provide valuable insights into various disease states such as inflammatory disorders, autoimmune diseases, and cardiovascular conditions, making this ELISA kit an essential resource for researchers in these fields.
Overall, the Mouse P-Selectin Glycoprotein Ligand-1 (PSGL-1) ELISA Kit offers a powerful tool for investigating the role of PSGL-1 in various physiological and pathological processes, enabling researchers to advance their understanding of this important protein and its potential implications in disease development and treatment strategies.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.