The Mouse Lactate Dehydrogenase A (LDHA) ELISA Kit is a specialized tool for the accurate quantification of LDHA levels in mouse serum, plasma, and cell culture samples. This kit offers exceptional sensitivity and specificity, ensuring precise and reproducible results for a variety of research applications.LDHA is a key enzyme involved in the process of converting lactate to pyruvate during anaerobic metabolism. Dysregulation of LDHA has been linked to various diseases and conditions, including cancer, metabolic disorders, and neurodegenerative diseases.
The measurement of LDHA levels can provide valuable insights into cellular metabolism and disease progression, making it an indispensable tool for researchers in the field.With the Mouse LDHA ELISA Kit, researchers can accurately assess LDHA levels in biological samples and gain a deeper understanding of its role in health and disease. This kit is easy to use, reliable, and suitable for high-throughput analysis, making it an essential resource for any laboratory studying LDHA function and its implications in various pathologies.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.