The Mouse IL-22 ELISA Kit is specifically designed for the precise quantification of interleukin-22 levels in mouse serum, plasma, and cell culture supernatants. This kit offers exceptional sensitivity and specificity, guaranteeing accurate and consistent results for various research applications.Interleukin-22 is a critical cytokine that plays a vital role in the regulation of immune response and inflammation. It is involved in various physiological processes, including tissue repair, antimicrobial defense, and mucosal immunity.
Elevated levels of IL-22 have been associated with various inflammatory conditions and autoimmune diseases, making it a valuable biomarker for studying these disorders and potential therapeutic interventions.Overall, the Mouse IL-22 ELISA Kit is a valuable tool for researchers looking to investigate the role of IL-22 in immune-related diseases and explore its potential as a target for novel therapeutic strategies.
Product Name:
Mouse IL-22 ELISA Kit
SKU:
MOES01725
Target:
Mouse IL-22
Size:
96T
Assay type:
Sandwich-ELISA
Assay time:
3.5h
Sensitivity:
9.38 pg/mL
Detection range:
15.63-1000 pg/mL
Kit component:
Component
Quantity (24 Assays)
Quantity (96 Assays)
Storage
Micro ELISA Plate (Dismountable)
8 wells x 3 strips
8 wells x 12 strips
-20°C, 12 months
Reference Standard
1 vial
2 vials
Concentrated Biotinylated Detection Ab (100×)
1 vial, 60 µL
1 vial, 120 µL
Concentrated HRP Conjugate (100×)
1 vial, 60 µL
1 vial, 120 µL
-20°C (shading light), 12 months
Reference Standard & Sample Diluent
1 vial, 20 mL
1 vial, 20 mL
4°C, 12 months
Biotinylated Detection Ab Diluent
1 vial, 14 mL
1 vial, 14 mL
HRP Conjugate Diluent
1 vial, 14 mL
1 vial, 14 mL
Concentrated Wash Buffer (25×)
1 vial, 30 mL
1 vial, 30 mL
Substrate Reagent
1 vial, 10 mL
1 vial, 10 mL
4°C (shading light)
Stop Solution
1 vial, 10 mL
1 vial, 10 mL
4°C
Plate Sealer
5 pieces
5 pieces
Product Description
1 copy
1 copy
This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Mouse IL-22. Standards or samples are added to the micro ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Mouse IL-22 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Mouse IL-22, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The OD value is proportional to the concentration of Mouse IL-22. You can calculate the concentration of Mouse IL-22 in the samples by comparing the OD of the samples to the standard curve.
Linearity:
Serum (n=5)
EDTA plasma (n=5)
Cell culture media (n=5)
1:2
Range (%)
97-115
92-107
95-110
Average (%)
105
98
103
1:4
Range (%)
88-101
83-97
81-95
Average (%)
94
88
87
1:8
Range (%)
91-107
83-96
83-99
Average (%)
99
90
90
1:16
Range (%)
86-98
79-93
84-98
Average (%)
93
85
91
Recovery:
Sample Type
Range (%)
Average Recovery (%)
Serum (n=5)
87-101
94
EDTA plasma (n=5)
91-104
96
Cell culture media (n=5)
84-98
91
Precision:
Intra-assay Precision
Inter-assay Precision
Sample
1
2
3
1
2
3
n
20
20.0
20.0
20.0
20.0
20.0
Mean (ng/mL)
48
111.4
386.5
49.14
119.2
389.27
Standard deviation
2.95
6.42
20.76
2.86
5.71
17.75
C V (%)
6.15
5.76
5.37
5.82
4.79
4.56
Sample type &Sample volume:
serum, plasma and other biological fluids; 100μL
Reproducibility:
Both intra-CV and inter-CV are < 10%.
Application:
This ELISA kit applies to the in vitro quantitative determination of Mouse IL-22 concentrations in serum, plasma and other biological fluids.
Specificity:
This kit recognizes Mouse IL-22 in samples. No significant cross-reactivity or interference between Mouse IL-22 and analogues was observed.