The Mouse Hyaluronoglucosaminidase-1 (HYAL1) ELISA Kit is a comprehensive solution for the detection of HYAL1 levels in mouse serum, plasma, and cell culture supernatants. This kit is highly sensitive and specific, providing accurate and consistent results for a variety of research applications.HYAL1 is an important enzyme that plays a key role in the breakdown of hyaluronic acid, a major component of the extracellular matrix. Dysregulation of HYAL1 has been linked to various diseases, including cancer, inflammation, and tissue remodeling.
Therefore, monitoring HYAL1 levels can offer valuable insights into disease progression and potential therapeutic strategies.With the Mouse HYAL1 ELISA Kit, researchers can confidently measure HYAL1 levels with precision and efficiency, enabling them to advance their understanding of disease mechanisms and develop targeted interventions. This kit is a valuable tool for investigating the intricate interplay between HYAL1 and disease pathways in mouse models.
LUCA1, NAT6, Hyaluronidase-1, Lung carcinoma protein 1
Detection Method:
Sandwich
Range:
78.125-5000pg/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.