The Mouse Copeptin (CPP) ELISA Kit is specifically designed for the precise measurement of copeptin levels in mouse serum, plasma, and cell culture supernatants. This kit offers exceptional sensitivity and specificity, guaranteeing accurate and consistent results, which makes it an excellent tool for various research purposes.Copeptin is a key peptide involved in the regulation of vasopressin release, which plays a critical role in fluid balance, blood pressure regulation, and stress response.
Elevated copeptin levels have been associated with various medical conditions, including cardiovascular diseases, kidney disorders, and sepsis, making it a valuable biomarker for studying these diseases and exploring potential therapeutic interventions.Overall, the Mouse Copeptin (CPP) ELISA Kit provides a reliable and efficient method for assessing copeptin levels in mouse samples, facilitating research in areas such as physiology, pharmacology, and disease mechanisms.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-99
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition. Note:To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity, and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 50µL standard or sample to each well.    And then add 50µL prepared Detection Reagent A immediately.    Shake and mix. Incubate 1 hour at 37°C
3.
Aspirate and wash 3 times
4.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
5.
Aspirate and wash 5 times
6.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
7.
Add 50µL Stop Solution. Read at 450 nm immediately.