The Mouse Calcium Channel Voltage-Dependent L-Type Alpha-1F Subunit (CACNA1F) ELISA Kit is a powerful tool for the accurate quantification of CACNA1F levels in mouse samples. This kit is highly sensitive and specific, ensuring precise and reproducible results in serum, plasma, and cell culture supernatants.CACNA1F is a key component of voltage-gated calcium channels, playing a critical role in calcium influx and neuronal excitability. Dysregulation of CACNA1F has been implicated in various neurological disorders, including autism spectrum disorders and X-linked congenital stationary night blindness.
By measuring CACNA1F levels, researchers can gain valuable insights into the mechanisms underlying these disorders and explore potential therapeutic targets.Whether you are studying neurobiology, ophthalmology, or other related fields, the Mouse Calcium Channel Voltage-Dependent L-Type Alpha-1F Subunit ELISA Kit is a valuable tool for advancing your research. Trust in its precision and reliability to support your scientific endeavors.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
83-100
91
EDTA plasma (n=5)
90-98
94
Heparin plasma (n=5)
89-102
95
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.