Mouse Breast And Kidney Expressed Chemokine (BRAK) ELISA Kit
The Mouse Breast and Kidney Expressed Chemokine (BRAK) ELISA Kit from AssayGenie is specially designed for the accurate and precise measurement of BRAK levels in mouse samples. This ELISA kit offers high sensitivity and specificity, ensuring dependable and consistent results for research purposes.BRAK, also known as CCL14, is a chemokine that plays a role in immune response and inflammation. It is expressed in various tissues, including the breast and kidney, and has been implicated in cancer, inflammatory diseases, and immune regulation.
The BRAK ELISA kit allows researchers to quantify BRAK levels in serum, plasma, and tissue culture supernatants, providing valuable insights into its biological functions and potential involvement in disease processes.With its user-friendly protocol and reliable performance, the Mouse Breast and Kidney Expressed Chemokine (BRAK) ELISA Kit is an essential tool for studying the role of BRAK in mouse models and advancing our understanding of its significance in health and disease.
Product Name:
Mouse Breast And Kidney Expressed Chemokine (BRAK) ELISA Kit
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.