The Monkey Transforming Growth Factor Beta 2 (TGFB2) ELISA Kit is a specialized assay designed for the accurate quantification of TGFB2 levels in monkey serum, plasma, and cell culture supernatants. This kit offers exceptional sensitivity and specificity, ensuring dependable and reproducible results for a variety of research applications.TGFB2 is a key growth factor that plays a critical role in cell growth, differentiation, and immune regulation. It is involved in various physiological processes such as wound healing, tissue development, and immune response modulation.
Dysregulation of TGFB2 has been implicated in numerous diseases, including cancer, fibrosis, and autoimmune disorders, highlighting its importance as a potential therapeutic target and diagnostic marker.By utilizing the Monkey TGFB2 ELISA Kit, researchers can accurately measure TGFB2 levels in biological samples, paving the way for a deeper understanding of its role in health and disease. This kit is a valuable tool for studying monkey models and advancing scientific knowledge in the fields of immunology, oncology, and developmental biology.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
83-99
91
EDTA plasma (n=5)
80-99
89
Heparin plasma (n=5)
82-99
90
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.