Kit Name: | MG (Malachite Green) ELISA Kit |
Kit Size: | 96 Assays |
Test Principle: | This kit uses Competitive-ELISA as the method. It can detect Malachite Green (MG) in samples, such as muscle, etc. This kit is composed of ELISA Microtiter plate, HRP conjugate, antibody working solution, standard and other supplementary reagents. The microtiter plate in this kit has been pre-coated with coupled antigen. During the reaction, MG in the samples or standard competes with coupled antigen on the solid phase supporter for sites of anti-MG antibody. Then Horseradish Peroxidase (HRP) conjugate is added to each microtiter plate well, and substrate reagent is added for color development. There is a negative correlation between the OD value of samples and the concentration of MG. The concentration of MG in the samples can be calculated by comparing the OD of the samples to the standard curve. |
Sensitivity: | 0.025 ppb (ng/mL) |
Reaction Mode: | 25°C, 30 min~30 min~15min |
Detection Limit: | Muscle ---0.1 ppb |
Cross Reactivity: | Malachite Green --100%, Crystal violet---80%, Leucomalachite green (oxidized) --- 100%, Leucocrystal Violet (oxidized) --- 80% |
Sample Recovery Rate: | Muscle --85%±15%. |
Kits Components: | Item | Specifications | ELISA Microtiter plate | 96 wells | High Concentrated Standard (10 ppb) | 1 mL | Standard Liquid (empty bottles) | (0 ppb, 0.025 ppb, 0.05 ppb, 0.1 ppb, 0.2 ppb, 0.4 ppb) | HRP Conjugate | 11 mL | Antibody Working Solution | 5.5 mL | Substrate Reagent A | 6 mL | Substrate Reagent B | 6 mL | Stop Solution | 6 mL | Cosolvent | 6 mL | Oxygenant | 6 mL | 20 x Concentrated Wash Buffer | 40 mL | 10 x Reconstitution Buffer | 20 mL | Plate Sealer | 3 pieces | Sealed Bag | 1 piece | Manual | 1 copy | |
Instruments: | Microplate reader, Printer, Homogenizer, Nitrogen evaporators, Water bath, Oscillators, Centrifuge, Graduated pipette, Balance (sensibility 0.01 g). |
Micropipette: | Single channel (20-200 uL, 100-1000 uL), Multichannel (30-300 uL). |
Reagents: | Acetonitrile, Ethyl acetate, Methanol. |
Storage And Expiry Date: | Store the kit at 2~8°C. Do not freeze any test kit components. Return any unused microwells to their original foil bag and reseal them together with the desiccant provided and further store at 2 - 8°C. Expiry date: expiration date is on the packing box. |