Human X-Ray Repair Cross Complementing 6 (XRCC6) ELISA Kit
SKU:
HUDL03148
UPC:
Product Type:
ELISA Kit
Size:
96 Assays
Uniprot:
P12956
Sensitivity:
0.112ng/mL
Range:
0.312-20ng/mL
ELISA Type:
Sandwich
Synonyms:
Ku70,CTC75,CTCBF,G22P1,ML8,TLAA,Lupus Ku Autoantigen Protein P70,ATP-Dependent DNA Helicase 2 Subunit 1,CTC box-binding factor 75 kDa,Thyroid-lupus autoantigen
Reactivity:
Human
Sample Type:
Tissue Homogenates, Cell Lysates And Other Biological Fluids
Human X-Ray Repair Cross Complementing 6 (XRCC6) ELISA Kit
The Human X-Ray Repair Cross Complementing 6 (XRCC6) ELISA Kit is a highly sensitive and specific assay designed for the detection of XRCC6 levels in human samples. XRCC6 is a key protein involved in DNA repair, particularly in response to DNA damage caused by exposure to ionizing radiation.This ELISA kit offers reliable and reproducible results, making it ideal for a wide range of research applications related to DNA repair mechanisms, radiation-induced damage, and genomic stability.
By accurately quantifying XRCC6 levels in serum, plasma, or cell culture supernatants, researchers can gain valuable insights into the molecular pathways involved in DNA repair processes.Overall, the Human XRCC6 ELISA Kit is an essential tool for studying the role of XRCC6 in maintaining genomic integrity and understanding its potential implications in diseases such as cancer and genetic disorders.
Product Name:
Human X-Ray Repair Cross Complementing 6 (XRCC6) ELISA Kit
Product Code:
HUDL03148
Size:
96 Assays
Target:
XRCC6
Synonyms:
Ku70 CTC75 CTCBF G22P1 ML8 TLAA Lupus Ku Autoantigen Protein P70 ATP-Dependent DNA Helicase 2 Subunit 1 CTC box-binding factor 75 kDa Thyroid-lupus autoantigen
Detection Method:
Sandwich
Range:
0.312-20ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.