Human Uroporphyrinogen Decarboxylase (UROD) ELISA Kit
The Human Uroporphyrinogen Decarboxylase (UROD) ELISA Kit is designed for the precise measurement of UROD levels in human serum, plasma, and cell culture supernatants. With its high sensitivity and specificity, this kit delivers accurate and reproducible results, making it a valuable tool for various research purposes.Uroporphyrinogen decarboxylase is an enzyme critical for the biosynthesis of heme, an essential component of hemoglobin. Dysregulation of UROD has been linked to porphyria, a group of rare genetic disorders characterized by abnormal heme production.
This ELISA kit enables the reliable quantification of UROD levels, allowing researchers to better understand the role of this enzyme in porphyria and other related conditions.Overall, the Human Uroporphyrinogen Decarboxylase (UROD) ELISA Kit is a reliable and efficient solution for studying UROD-related disorders and advancing research in the field of heme biosynthesis.
Product Name:
Human Uroporphyrinogen Decarboxylase (UROD) ELISA Kit
Product Code:
HUDL03061
Size:
96 Assays
Target:
UROD
Synonyms:
PCT, Uroporphyrinogen III Decarboxylase
Detection Method:
Sandwich
Range:
0.156-10ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.