Human Ubiquitin Specific Peptidase 14 (USP14) ELISA Kit
The Human Ubiquitin-Specific Peptidase 14 (USP14) ELISA Kit is a powerful tool for the quantitative measurement of USP14 levels in human biological samples. This kit offers exceptional sensitivity and specificity, providing researchers with accurate and reliable results for various research applications.USP14 is a key enzyme involved in the regulation of protein degradation pathways, specifically in the ubiquitin-proteasome system.
Dysregulation of USP14 has been implicated in various diseases, including neurodegenerative disorders and cancer. Therefore, analyzing USP14 levels can provide valuable insights into disease mechanisms and potential therapeutic targets.Overall, the Human USP14 ELISA Kit from AssayGenie is a valuable tool for researchers looking to study the role of USP14 in disease progression and therapeutic interventions.
Product Name:
Human Ubiquitin Specific Peptidase 14 (USP14) ELISA Kit
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
83-100
91
EDTA plasma (n=5)
90-103
96
Heparin plasma (n=5)
89-102
95
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.