Human Tissue Inhibitors Of Metalloproteinase 4 (TIMP4) ELISA Kit
The Human TIMP-4 (Tissue Inhibitors of Metalloproteinase-4) ELISA Kit is specially developed for the precise measurement of TIMP-4 levels in human serum, plasma, and cell culture supernatants. With its high sensitivity and specificity, this kit delivers accurate and consistent results, making it an excellent choice for a variety of research purposes.TIMP-4 is a key regulator of metalloproteinases, enzymes that play a crucial role in tissue remodeling and cell migration.
Dysregulation of TIMP-4 has been implicated in various diseases, including cancer, cardiovascular diseases, and arthritis. By accurately measuring TIMP-4 levels, researchers can gain valuable insights into the pathogenesis of these conditions and potentially identify new therapeutic targets.
Product Name:
Human Tissue Inhibitors Of Metalloproteinase 4 (TIMP4) ELISA Kit
Product Code:
HUDL02900
Size:
96 Assays
Target:
TIMP4
Synonyms:
TIMP Metallopeptidase Inhibitor 4, Tissue Inhibitor of Metalloproteinases 4
Detection Method:
Sandwich
Range:
0.156-10ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.