The Human Thyroxine Binding Globulin (TBG) ELISA Kit is specifically designed for the precise and reliable measurement of TBG levels in human serum, plasma, and cell culture supernatants. With its high sensitivity and specificity, this kit ensures accurate and reproducible results, making it an invaluable tool for various research applications.Thyroxine Binding Globulin (TBG) is a crucial protein that plays a vital role in the transport of thyroid hormones in the bloodstream. Dysregulation of TBG levels can lead to hormonal imbalances and thyroid disorders, making it a key biomarker for studying thyroid function and related conditions.
Whether studying thyroid diseases, hormonal imbalances, or exploring potential therapeutic interventions, the Human TBG ELISA Kit provides researchers with a reliable and efficient method for measuring TBG levels in a variety of biological samples. Trust in the accuracy and precision of this ELISA kit to advance your research and uncover new insights into thyroid physiology and pathology.
Product Name:
Human Thyroxine Binding Globulin (TBG) ELISA Kit
Product Code:
HUDL02837
Size:
96 Assays
Target:
TBG
Synonyms:
SERPINA7, Serpin Peptidase Inhibitor Clade A Member 7, T4-binding globulin
Detection Method:
Sandwich
Range:
0.781-50ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-99
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.