Human Suppression Of Tumorigenicity 14 (ST14) ELISA Kit
The Human Suppression of Tumorigenicity 14 (ST14) ELISA Kit is specifically designed for the accurate measurement of ST14 levels in human samples such as serum, plasma, and cell culture supernatants. This kit provides high sensitivity and specificity, ensuring precise and reliable results for various research applications.ST14, also known as matriptase, is a transmembrane serine protease that plays a critical role in cell growth, invasion, and metastasis in cancer. Elevated levels of ST14 have been associated with various types of cancer, making it a valuable biomarker for studying tumorigenesis and developing potential therapeutic strategies.
By utilizing the Human ST14 ELISA Kit, researchers can gain valuable insights into the role of ST14 in cancer progression and potentially identify new targets for cancer treatment. With its ease of use and robust performance, this kit is an essential tool for advancing cancer research and improving patient outcomes.
Product Name:
Human Suppression Of Tumorigenicity 14 (ST14) ELISA Kit
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
83-100
91
EDTA plasma (n=5)
90-103
96
Heparin plasma (n=5)
89-102
95
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.