Human Superoxide Dismutase 1, Soluble (SOD1) ELISA Kit
The Human Superoxide Dismutase 1 (SOD1) ELISA Kit is a powerful tool designed for the precise measurement of soluble SOD1 levels in human samples such as serum, plasma, and cell culture supernatants. With its exceptional sensitivity and specificity, this kit provides accurate and consistent results, making it an invaluable resource for various research purposes.SOD1 is an important antioxidant enzyme that plays a crucial role in combating oxidative stress and protecting cells from damage. Dysregulation of SOD1 has been linked to various diseases including neurodegenerative disorders, cardiovascular diseases, and cancer.
As such, measuring SOD1 levels can provide valuable insights into disease progression and aid in the development of potential treatments.Whether you are studying oxidative stress, neurodegeneration, or other related areas of research, the Human Superoxide Dismutase 1 ELISA Kit offers a reliable and efficient solution for detecting and quantifying soluble SOD1 levels in human samples. Elevate your research with this cutting-edge kit today.
Product Name:
Human Superoxide Dismutase 1, Soluble (SOD1) ELISA Kit
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.