Human Secondary Lymphoid Tissue Chemokine (SLC) ELISA Kit
The Human Secondary Lymphoid Tissue Chemokine (SLC) ELISA Kit is designed for precise measurement of SLC levels in human samples such as serum, plasma, and cell culture supernatants. This kit offers superior sensitivity and specificity, guaranteeing accurate and consistent results for various research applications.SLC, also known as CCL21, is a chemokine crucial for immune cell trafficking and lymphoid tissue organization. It plays a vital role in immune responses, inflammation, and autoimmune diseases.
The detection of SLC levels can provide valuable insights into immune system function and disease pathogenesis.This ELISA kit provides a reliable tool for studying the role of SLC in immune regulation and disease development, facilitating the discovery of potential therapeutic targets and strategies. With its high performance and user-friendly format, the Human Secondary Lymphoid Tissue Chemokine ELISA Kit is an essential asset for immunology and inflammation research.
Product Name:
Human Secondary Lymphoid Tissue Chemokine (SLC) ELISA Kit
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-99
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.