Human RNA Binding Motif Protein 38 (RBM38) ELISA Kit
The Human RNA Binding Motif Protein 38 (RBM38) ELISA Kit is specifically designed for the precise measurement of RBM38 levels in human samples including serum, plasma, and cell culture supernatants. This ELISA kit offers exceptional sensitivity and specificity, ensuring accurate and consistent results for various research applications.RBM38 is a key RNA-binding protein that plays a critical role in post-transcriptional regulation of gene expression. It is involved in regulating mRNA stability and translation, making it a crucial player in various cellular processes.
Dysregulation of RBM38 has been linked to various diseases including cancer, cardiovascular disorders, and neurological conditions, highlighting its potential as a valuable biomarker for disease research and therapeutic development.With its reliable performance and user-friendly format, the RBM38 ELISA Kit from Assay Genie provides researchers with a valuable tool for studying the role of RBM38 in disease pathogenesis and potentially identifying novel therapeutic targets.
Product Name:
Human RNA Binding Motif Protein 38 (RBM38) ELISA Kit
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
83-100
91
EDTA plasma (n=5)
91-98
94
Heparin plasma (n=5)
90-99
94
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.