Human Phosphodiesterase 1A, Calmodulin Dependent (PDE1A) ELISA Kit
The Human Phosphodiesterase 1A (Calmodulin-dependent PDE1A) ELISA Kit is a powerful tool for the accurate measurement of PDE1A levels in human samples such as serum, plasma, and cell culture supernatants. This kit offers exceptional sensitivity and specificity, ensuring precise and consistent results for a variety of research applications.PDE1A is a critical enzyme involved in the regulation of intracellular levels of cyclic nucleotides, playing a vital role in various physiological processes such as cell signaling, smooth muscle contraction, and neuronal function.
Dysregulation of PDE1A has been implicated in a range of diseases including cardiovascular disorders, neurological conditions, and cancer, making it a valuable biomarker for understanding disease mechanisms and developing potential therapeutic interventions.With its high-performance capabilities and ease of use, the Human Phosphodiesterase 1A (Calmodulin-dependent PDE1A) ELISA Kit is an essential tool for researchers seeking to explore the role of PDE1A in health and disease.
Product Name:
Human Phosphodiesterase 1A, Calmodulin Dependent (PDE1A) ELISA Kit
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
83-100
91
EDTA plasma (n=5)
90-103
96
Heparin plasma (n=5)
89-102
95
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.