The Human Peroxiredoxin-1 (PRDX1) ELISA Kit is a high-quality assay designed for the precise measurement of PRDX1 levels in human serum, plasma, and cell culture supernatants. With its superior sensitivity and specificity, this kit ensures accurate and consistent results, making it an excellent choice for various research applications.PRDX1 is a critical antioxidant enzyme that plays a crucial role in protecting cells from oxidative stress and maintaining cellular homeostasis. Dysregulation of PRDX1 has been implicated in various diseases, including cancer, neurodegenerative disorders, and cardiovascular diseases, highlighting its significance as a potential biomarker for disease diagnosis and prognosis.
By using the Human PRDX1 ELISA Kit, researchers can effectively study the role of PRDX1 in disease pathogenesis, assess its potential as a therapeutic target, and monitor its levels in response to treatment interventions. This kit provides a reliable tool for advancing our understanding of PRDX1 biology and its implications for human health.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.