The Human Nucleoporin 62kDa (NUP62) ELISA Kit is designed for the accurate detection of NUP62 levels in human serum, plasma, and cell culture supernatants. This kit features high sensitivity and specificity, ensuring reliable and reproducible results for researchers studying nucleoporin functions.Nucleoporin 62kDa (NUP62) is a key component of the nuclear pore complex, playing a critical role in regulating nucleocytoplasmic transport and maintaining cellular homeostasis.
Dysregulation of NUP62 has been associated with various diseases, including cancer and autoimmune disorders, making it a valuable target for therapeutic research.Overall, the Human Nucleoporin 62kDa (NUP62) ELISA Kit provides a valuable tool for investigating the roles of NUP62 in cellular processes and disease pathogenesis, offering researchers a reliable platform for biomarker analysis and drug development studies.
Product Name:
Human Nucleoporin 62kDa (NUP62) ELISA Kit
Product Code:
HUDL02118
Size:
96 Assays
Target:
NUP62
Synonyms:
IBSN, p62, SNDI, Nuclear Pore Glycoprotein p62
Detection Method:
Sandwich
Range:
0.312-20ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
83-100
91
EDTA plasma (n=5)
90-99
94
Heparin plasma (n=5)
91-101
96
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.