Human Nitric Oxide Synthase 2, Inducible (NOS2) ELISA Kit
The Human Nitric Oxide Synthase 2 (Inducible NOS2) ELISA Kit is a cutting-edge tool for the precise measurement of NOS2 levels in human samples such as serum, plasma, and cell culture supernatants. With its exceptional sensitivity and specificity, this kit delivers accurate and consistent results, making it an invaluable asset for various research endeavors. NOS2 is a critical enzyme responsible for the production of nitric oxide, a key signaling molecule involved in a range of physiological processes, including immune response and inflammation.
Dysregulation of NOS2 has been implicated in numerous diseases, such as inflammatory disorders, cardiovascular conditions, and neurodegenerative illnesses, underscoring the importance of studying its levels as a potential biomarker for disease progression and therapeutic interventions.Incorporate the Human Nitric Oxide Synthase 2 (Inducible NOS2) ELISA Kit into your research toolkit and unlock the potential for groundbreaking discoveries in the field of nitric oxide biology and related pathologies.
Product Name:
Human Nitric Oxide Synthase 2, Inducible (NOS2) ELISA Kit
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.