Human Nitric Oxide Synthase 1, Neuronal (NOS1) ELISA Kit
The Human Neuronal Nitric Oxide Synthase 1 (NOS1) ELISA Kit is a powerful tool for the accurate measurement of NOS1 levels in human samples such as serum, plasma, and cell culture supernatants. With its high sensitivity and specificity, this kit delivers reliable and reproducible results, making it ideal for a variety of research applications.Neuronal Nitric Oxide Synthase 1 (NOS1) is an important enzyme involved in the production of nitric oxide, a key signaling molecule in the body. NOS1 is known to play a critical role in various physiological processes, including neurotransmission, smooth muscle relaxation, and immune response.
Dysregulation of NOS1 has been linked to a range of diseases, such as neurodegenerative disorders, cardiovascular diseases, and inflammatory conditions, making it a valuable target for research and potential therapeutic interventions.The Human Neuronal Nitric Oxide Synthase 1 (NOS1) ELISA Kit offers researchers a reliable and accurate method to quantify NOS1 levels in human samples, providing valuable insights into the role of this enzyme in health and disease. Order your kit today and take your research to the next level.
Product Name:
Human Nitric Oxide Synthase 1, Neuronal (NOS1) ELISA Kit
Product Code:
HUDL02056
Size:
96 Assays
Target:
NOS1
Synonyms:
NOS, nNOS, IHPS1, NC-NOS, N-NOS, bNOS, Neuronal NOS, Constitutive NOS, NOS type I, Nitric Ocide Synthase, Peptidyl-cysteine S-nitrosylase NOS1
Detection Method:
Sandwich
Range:
0.156-10ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.