The Human NELL2 (Nel-like protein 2) ELISA Kit is specifically designed for the accurate quantification of NELL2 levels in human serum, plasma, and cell culture supernatants. This kit boasts high sensitivity and specificity, ensuring precise and reproducible results, making it an excellent choice for a variety of research applications.NELL2 is a vital protein involved in various biological processes, including neuronal development, cell adhesion, and synaptic plasticity. Dysregulation of NELL2 has been linked to conditions such as neurodevelopmental disorders, neurodegenerative diseases, and cancer, highlighting its importance as a potential biomarker for studying these conditions and developing targeted therapies.
With the Human NELL2 ELISA Kit, researchers can accurately measure NELL2 levels in biological samples, providing valuable insights into its role in health and disease. This kit is user-friendly, reliable, and efficient, making it an essential tool for advancing research in the field of neuroscience and related disciplines.
Product Name:
Human NEL Like Protein 2 (NELL2) ELISA Kit
Product Code:
HUDL02009
Size:
96 Assays
Target:
NELL2
Synonyms:
NRP2, Nel-related protein 2, Protein kinase C-binding protein NELL2
Detection Method:
Sandwich
Range:
0.312-20ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.