The Human NADH Dehydrogenase 1 (ND1) ELISA Kit is specifically designed for the precise measurement of ND1 levels in human samples including serum, plasma, and cell culture supernatants. With its exceptional sensitivity and specificity, this kit delivers consistent and accurate results, making it a valuable tool for a multitude of research applications.ND1 is a key enzyme involved in the electron transport chain and energy production within mitochondria. Dysregulation of ND1 has been linked to various diseases such as mitochondrial disorders, neurodegenerative diseases, and aging-related conditions.
As such, the measurement of ND1 levels can provide valuable insights into these pathologies and aid in the development of potential therapeutic interventions.With the Human NADH Dehydrogenase 1 (ND1) ELISA Kit, researchers can obtain reliable data to further their understanding of mitochondrial function and its implications in health and disease. This kit offers excellent performance, convenience, and accuracy, making it an indispensable tool for studying ND1 biology and its related disorders.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-99
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.