The Human Myosin VIIB (MYO7B) ELISA Kit is a cutting-edge tool designed for the precise measurement of MYO7B levels in human samples including serum, plasma, and cell culture supernatants. This advanced kit offers exceptional sensitivity and specificity, ensuring accurate and consistent results for a variety of research applications.MYO7B is a critical protein involved in cell motility and function, particularly in the maintenance of cellular structures and intracellular transport. Dysregulation of MYO7B has been linked to various diseases such as certain types of deafness and kidney disorders, highlighting its importance as a potential biomarker for disease diagnostics and therapeutic development.
With its reliable performance and versatility, the Human Myosin VIIB (MYO7B) ELISA Kit is an invaluable tool for researchers seeking to explore the role of MYO7B in health and disease, advancing our understanding of cellular mechanisms and potentially leading to novel treatment strategies.
Product Name:
Human Myosin VIIB (MYO7B) ELISA Kit
Product Code:
HUDL01968
Size:
96 Assays
Target:
MYO7B
Detection Method:
Sandwich
Range:
0.312-20ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
83-100
91
EDTA plasma (n=5)
91-100
95
Heparin plasma (n=5)
91-101
96
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.