Human Minichromosome Maintenance Deficient 5 (MCM5) ELISA Kit
The Human Minichromosome Maintenance Deficient 5 (MCM5) ELISA Kit is a valuable tool for the accurate measurement of MCM5 levels in human serum, plasma, and cell culture supernatants. This kit boasts exceptional sensitivity and specificity, ensuring precise and consistent results that are essential for a variety of research applications.MCM5 is a key protein involved in DNA replication and cell cycle regulation, playing a critical role in maintaining genomic integrity.
Dysregulation of MCM5 has been linked to cancer and other diseases, making it a promising biomarker for studying these conditions and developing new treatment strategies.Overall, the Human MCM5 ELISA Kit offers researchers a reliable and efficient solution for studying the role of MCM5 in health and disease, providing valuable insights into cellular processes and potential therapeutic targets.
Product Name:
Human Minichromosome Maintenance Deficient 5 (MCM5) ELISA Kit
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.