The Human Meprin A Beta (MEP1B) ELISA Kit is a powerful tool designed for the precise measurement of MEP1B levels in human serum, plasma, and cell culture supernatants. This kit boasts exceptional sensitivity and specificity, guaranteeing accurate and consistent results for a variety of research purposes.MEP1B, also known as Meprin A Subunit Beta, is a key enzyme involved in the regulation of extracellular matrix remodeling and inflammation. It plays a crucial role in various biological processes and has been implicated in diseases such as inflammatory bowel disease, fibrosis, and cancer.
Detecting and quantifying MEP1B levels can provide valuable insights into these conditions and aid in the development of potential therapeutic interventions.The Human MEP1B ELISA Kit from Assay Genie is a reliable and efficient tool for researchers seeking to investigate the role of MEP1B in health and disease. With easy-to-follow protocols and rapid results, this kit is an essential addition to any laboratory conducting research on enzyme function and disease pathogenesis.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
83-100
91
EDTA plasma (n=5)
90-103
96
Heparin plasma (n=5)
89-102
95
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.